Model Number 112266-001
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/17/2016
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.(b)(4).Device returned to mfr.:returned product consisted of a jetstream xc-2.4 atherectomy catheter.The catheter shaft, pod, and the rest of the device was checked for damage.Product analysis tested the device by connecting it to the jetstream console.The functionality of the device was consistent with the design.The device ran as intended with no issues.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
 
Event Description
It was reported the device stopped spinning.A jetstream® xc atherectomy catheter was selected for an in stent restenosis atherectomy procedure in the superficial femoral artery.After the second pass through the lesion when the blades were up, the device stopped spinning after three quarters of the way through.This occurred when the device was in a fibrous portion of the lesion between the stents.They took it out of the body and flushed the catheter.It started working again and was placed back into the body.However, it stopped working after three seconds.The procedure was completed with another of the same device and there were nice results.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JETSTREAM® XC ATHERECTOMY CATHETER
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key6099103
MDR Text Key59752856
Report Number2134265-2016-09942
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130637
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 10/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2018
Device Model Number112266-001
Device Catalogue NumberPV41340
Device Lot Number0019622678
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/17/2016
Initial Date FDA Received11/14/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/31/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-