Model Number MODEL 250 |
Device Problems
Mechanical Problem (1384); Programming Issue (3014)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/20/2016 |
Event Type
malfunction
|
Event Description
|
It was reported that a physician's programming system was having issues.The physician was able to use a back up system, but hers was not operational.The company representative who reported the issue stated that the programming system's serial cable was the issue.No further relevant information has been received to date.
|
|
Event Description
|
Troubleshooting was performed to narrow the cause of the programming issues to the tablet serial cable.The serial cable was unplugged and plugged back in multiple times with no success.After a known good cable was used, the issues stopped, indicating that the interrogation difficulties were due to the serial cable.
|
|
Event Description
|
Further information was provided verifying that the serial cable was discarded after it was identified to be the cause of the issue.No device evaluation is needed because the failure of the serial cable has a known and well understood failure mode.
|
|