Model Number MODEL 250
Device Problems Mechanical Problem (1384); Programming Issue (3014)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/20/2016
Event Type  malfunction  
Event Description
It was reported that a physician's programming system was having issues.The physician was able to use a back up system, but hers was not operational.The company representative who reported the issue stated that the programming system's serial cable was the issue.No further relevant information has been received to date.
 
Event Description
Troubleshooting was performed to narrow the cause of the programming issues to the tablet serial cable.The serial cable was unplugged and plugged back in multiple times with no success.After a known good cable was used, the issues stopped, indicating that the interrogation difficulties were due to the serial cable.
 
Event Description
Further information was provided verifying that the serial cable was discarded after it was identified to be the cause of the issue.No device evaluation is needed because the failure of the serial cable has a known and well understood failure mode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6100209
MDR Text Key60312581
Report Number1644487-2016-02636
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 10/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 11.0.5
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Other
Initial Date Manufacturer Received 10/20/2016
Initial Date FDA Received11/14/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received12/05/2016
12/19/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/27/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-