• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT22719
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/16/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(6) 2016 that on (b)(6) 2016, the receiver had no display except the backlight.No additional event or patient information is available.The complaint device was returned for evaluation.The device was visually inspected and no defect was found.Functional testing was performed and there was no failure detected related to the complaint.The receiver log was reviewed and firmware errors, hardware errors, and screen error alarms were observed.The reported event of no receiver display except backlight was confirmed during log review.The root cause was determined to be a bad receiver battery.A root cause could not be determined.Based on the investigation results this issue has been upgraded from non-reportable to reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key6100942
MDR Text Key59888713
Report Number3004753838-2016-28662
Device Sequence Number1
Product Code MDS
UDI-Device Identifier00386270000231
UDI-Public(01)00386270000231(241)MT22719(10)5204546(17)NA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 10/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT22719
Device Catalogue NumberSTK-GF-001
Device Lot Number5204546
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/04/2016
Initial Date FDA Received11/14/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/10/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age16 YR
Patient Weight45
-
-