Lot Number E207-KIT
Device Problem Crack (1135)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/17/2016
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.A batch record review of kit lot e207 was conducted.There were no nonconformances associated with this lot.The lot met release requirements.The (b)(4) lot number was not provided as it was not administered.However, a review of all (b)(4) lots manufactured since january 2013 was performed.No trends or nonconformances related to the complaint were noted.Trends were reviewed for complaint category, tubing leak.No trends were detected for this complaint category.Service order report, #(b)(4), feedback: the service technician cleaned the instrument and replaced one pump head.The service technician then callibrated all the pumps and successfully completed the system checkout procedure.The assessment is based on information available at the time of the investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.(b)(4).Device not returned to manufacturer.
 
Event Description
The customer reported that the collect pump tubing segment "cracked" after 1300ml of whole blood processed.The customer stated that the treatment was aborted with no blood/products returned to the patient.The customer reported that the patient was in stable condition.Service was requested.The kit was not returned for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERAKOS CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX
Manufacturer (Section D)
THERAKOS, INC
hampton NJ
Manufacturer (Section G)
THERAKOS, INC.
10 north high street
suite 300
west chester PA 19380
Manufacturer Contact
megan vernak
53 frontage road
hampton, NJ 08827
MDR Report Key6101052
MDR Text Key60429546
Report Number2523595-2016-00252
Device Sequence Number1
Product Code LNR
UDI-Device Identifier10705030100009
UDI-Public(01)10705030100009(11)151007
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Nurse
Type of Report Initial
Report Date 10/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Expiration Date08/01/2018
Device Lot NumberE207-KIT
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/17/2016
Initial Date FDA Received11/14/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Weight80
-
-