STRYKER ORTHOPAEDICS-MAHWAH EXETER V40 STEM 44MM NO 1; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
|
Back to Search Results |
|
Catalog Number 0580-1-441 |
Device Problems
Break (1069); Crack (1135); Material Integrity Problem (2978)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
|
Event Date 10/18/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of the investigation.
|
|
Event Description
|
It was reported that, during a tha, the a centralizer was broken from the beginning.A spare was used instead.
|
|
Manufacturer Narrative
|
Reported event: an event regarding crack/fracture involving an exeter centraliser was reported.Crack/fracture of the exeter centraliser was confirmed following return of the device.Method and results: device evaluation and results: visual inspection of the returned device was carried out.The wingless centralizer shows one crack in the plastic, of a length of about 4mm.Medical records received and evaluation: no medical records were received for review with a clinical consultant.No adverse consequences to the patient were noted.Device history review: review indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other events for the lot referenced.Conclusions: nc has been raised to investigate the event.The scope of the non-conformance is one unit.Although the unit escaped, it was not implanted and has been returned through the complaint handling process and is now contained.There were no adverse consequences or delays to surgery reported in the surgery.As there are no other devices affected, there are no potential hazards or harms.
|
|
Event Description
|
It was reported that, during a tha, the a centralizer was broken from the beginning.A spare was used instead.
|
|