BIOMET MICROFIXATION ELEVATOR #301 Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The warnings in the package insert state this type of event can occur.Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event Description
It was reported the tip of the #301 elevator broke while removing tooth #17.The broken tip was retrieved via suction and another #301 elevator was used to complete the procedure.There was no injury to the patient.
 
Manufacturer Narrative
A visual inspection revealed moderate signs of wear including minor scratches along the length of the instrument along with a fractured tip.The device history record was reviewed and no non-conformances were identified.There are no indications of a manufacturing defect.The most likely underlying cause of the complaint was excessive force.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEVATOR #301
Type of Device
ELEVATOR
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer (Section G)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer Contact
michelle cole
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key6103284
MDR Text Key60027286
Report Number0001032347-2016-00675
Device Sequence Number1
Product Code EMJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PN/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number09-0257
Device Lot Number011014A14
Other Device ID NumberSEE H10 NARRATIVE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/27/2016
Initial Date FDA Received11/15/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/28/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/10/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction NumberN/A
Patient Sequence Number1
-
-