MIDMARK CORPORATION JB-70 INTRAORAL DENTAL X-RAY SYSTEM; UNIT, X-RAY, EXTRAORAL WITH TIMER Back to Search Results
Model Number JB-70
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Injury (2348)
Event Date 10/07/2016
Event Type  Injury  
Manufacturer Narrative
Inspection of the unit by an on-site service technician revealed that it was mounted on the wall with a single stud mount, with a lag screw into two sandwiched 3/4" press boards, the press board was mounted onto one 2x4 stud.The dry wall and press board (particle board) appeared damaged and there was no midmark installation guide tool used for this installation.The unit had been moved from its original location and installed in june or july 2016 by a third party company (exact date is unknown).The installer did not comply with the installation requirements provided in the jb-70 installation and service manual.The jb-70 installation and service manual details a proper procedure for installation of the unit which includes examples of using improper surfaces such as a particle board, and the need for an installation guide tool.Root cause of the mounting failure is determined to be improper installation by a third party company technician not certified by midmark.Should additional relevant information become available, a supplemental report will be submitted.Inspected on-site.
 
Event Description
It was reported that an x-ray unit had partially separated from the wall injuring a patient.The patient was struck by the tubehead and sustained injury to his nose and lip; patient was taken to urgent care.The reporter stated that the articulating arm also contacted the patient's chest however no injury resulted in the chest area.At the time of the incident the patient and dental assistant were present.The patient's recovery status and outcome of the event were not reported.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JB-70 INTRAORAL DENTAL X-RAY SYSTEM
Type of Device
UNIT, X-RAY, EXTRAORAL WITH TIMER
Manufacturer (Section D)
MIDMARK CORPORATION
675 heathrow drive
lincolnshire IL 60069
Manufacturer (Section G)
MIDMARK CORPORATION
675 heathrow drive
lincolnshire IL 60069
Manufacturer Contact
adam foresman
675 heathrow drive
lincolnshire, IL 60069
2242204329
MDR Report Key6104384
MDR Text Key59967879
Report Number1423380-2016-00001
Device Sequence Number1
Product Code EHD
Combination Product (y/n)N
PMA/PMN Number
K020070
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Service Personnel
Remedial Action Inspection
Type of Report Initial
Report Date 10/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Dental Assistant
Device Model NumberJB-70
Device Catalogue NumberI7017
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/18/2016
Initial Date FDA Received11/15/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/18/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-