• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY ICL (IMPLANTABLE COLLAMER LENS); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY ICL (IMPLANTABLE COLLAMER LENS); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VICMO12.6
Device Problems Device Operates Differently Than Expected (2913); Activation Failure (3270)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/08/2016
Event Type  malfunction  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.(b)(4).
 
Event Description
The reporter indicated the surgeon attempted to implant a 12.6mm vicmo12.6 implantable collamer lens, -11.0 diopter, in the patient's left eye (os) on (b)(6) 2016.The surgeon noticed the lens was folded and did not implant the lens.The backup lens was implanted.The patient's post-op best corrected visual acuity was 20/20.
 
Manufacturer Narrative
The lens was returned in a lens case/vial with clear surgical reside/debris on product.Visual inspection found the haptic bent and foreign material on lens surface (small particulates).(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICL (IMPLANTABLE COLLAMER LENS)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
michelle andres
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key6109400
MDR Text Key60231443
Report Number2023826-2016-01645
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date10/31/2018
Device Model NumberVICMO12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/25/2016
Initial Date FDA Received11/17/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/25/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/03/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age23 YR
-
-