SYNTHES OBERDORF SMALL BATTERY DRIVE (COLIBRI); INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.101 |
Device Problems
Defective Device (2588); Naturally Worn (2988)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/10/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.It was determined that due to lack of proper oiling the handpiece had a seized control lever and lock slides.The assignable root cause was determined to be due to improper maintenance, which is user error, misuse, and / or abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported from (b)(6) that it was hard to regulate the on/off button on the small battery drive device.During the pre-repair diagnostics assessment, it was determined that the control unit was not functioning and was defective.It was determined that the control lever and lock slides seized, the motor and bearings were worn out, and the lock slides were drilled out.It was further determined that the controller was defective.It was further determined that the device failed for check the off/oscillation/on switch mode function.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|