• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC® SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM-MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC® SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM-MAGNETIC ACTUATION Back to Search Results
Model Number RA002-4545SL70, RA002-4545SLR70
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem Therapeutic Response, Decreased (2271)
Event Type  Injury  
Manufacturer Narrative
The patient was implanted with dual magec rods (lot# a140919-05-00; manufacture date 09/01/2014, expiration date 09/01/2016, and lot# a140812-01-00; manufacture date 08/01/2014, expiration date 08/01/2016), and it was alleged that the rods appear to not be distracting.The rods were removed on (b)(6) 2016, and the patient was implanted with new dual magec rods without incident.To date, the patient is doing fine and no negative outcomes have been reported.A dhr review revealed that the rods met all of the required quality inspections and were released within specifications.
 
Event Description
A distributor reported that a surgeon alleged that a patient's dual magec rods appear to not be distracting after two (2) years of implantation.
 
Manufacturer Narrative
This supplemental report was sent to update the mfr report number.Supplemental 3001679046-2016-00039-1 was submitted under the incorrect report number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC® SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM-MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise
suite 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise
suite 100
aliso viejo CA 92656
Manufacturer Contact
john mcintyre
101 enterprise
suite 100
aliso viejo, CA 92656
9498373600
MDR Report Key6110231
MDR Text Key60252363
Report Number3006179046-2016-00039
Device Sequence Number1
Product Code PGN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140178
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model NumberRA002-4545SL70, RA002-4545SLR70
Device Lot NumberA140919-05-00, A140812-01-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/24/2016
Initial Date FDA Received11/17/2016
Supplement Dates Manufacturer Received01/06/2020
Supplement Dates FDA Received01/06/2020
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age8 YR
-
-