Model Number 103
Device Problems High impedance (1291); Malposition of Device (2616)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/25/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Information was received which indicates that a patient had high impedance seen during a system diagnostic test over a month after the patient had undergone a replacement surgery.X-rays were taken and reviewed.It was discovered that the pin was not fully inserted into the connector block.This is the most likely cause of the high impedance that was seen.During the patient's surgery on (b)(6) 2016 high impedance of 5668 ohms was seen but the pin was re-inserted and the next diagnostic test showed 4227 ohms.The lead pin was re-inserted again and the impedance was 3759 ohms.The pin was secured at that point and the patient was closed.Revision surgery has not occurred to date.No other relevant information has been received to date.
 
Event Description
Revision surgery occurred on (b)(6) 2016 where the lead pin was reinserted into the connector block and the high impedance resolved.No other relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 103
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6113815
MDR Text Key60860380
Report Number1644487-2016-02651
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 10/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/27/2017
Device Model Number103
Device Lot Number203335
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 10/25/2016
Initial Date FDA Received11/18/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/09/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/03/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age39 YR
-
-