• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS - HOUSTON PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS - HOUSTON PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problems Break (1069); Device Stops Intermittently (1599); Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/07/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the physician was having intermittent issues communicating with his programming system.A company representative performed troubleshooting to isolate the white serial cable as the issue.The cable would only work when held in by the company representative.After the serial cable was replaced, the communication issue immediately resolved.The broken serial cable was discarded.The serial cable does not require return, as the failure mode is understood to be a failure of the serial cable associated with a disconnected wire connection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6118952
MDR Text Key60878282
Report Number1644487-2016-02678
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 10/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 11.0.4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 10/28/2016
Initial Date FDA Received11/21/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/29/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-