It was reported that the physician was having intermittent issues communicating with his programming system.A company representative performed troubleshooting to isolate the white serial cable as the issue.The cable would only work when held in by the company representative.After the serial cable was replaced, the communication issue immediately resolved.The broken serial cable was discarded.The serial cable does not require return, as the failure mode is understood to be a failure of the serial cable associated with a disconnected wire connection.
|