• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. INDIGO SYSTEM ASPIRATION CATHETER 6; DXE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. INDIGO SYSTEM ASPIRATION CATHETER 6; DXE Back to Search Results
Catalog Number CAT6
Device Problems Kinked (1339); Physical Resistance (2578); Device Handling Problem (3265)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/22/2016
Event Type  malfunction  
Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2016-01661.The hospital disposed of the device.
 
Event Description
The patient was undergoing a thrombectomy procedure using indigo system aspiration catheter 6 (cat6) devices.During the procedure, while attempting to insert a cat6 into a non-penumbra sheath, the physician encountered resistance after the cat6 advanced pass the sheath¿s valve and subsequently, the cat6 became kinked.Therefore, the cat6 did not enter the sheath and was removed.The physician then attempted to insert a new cat6 into the same non-penumbra sheath.However, resistance was encountered again after the cat6 advanced pass the sheath's valve and subsequently, the cat6 became kinked.Therefore, the cat6 did not enter the sheath and was removed.It should be noted that the valve was too tight.The procedure was then completed using a new indigo system cat5 aspiration catheter (cat5).There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM ASPIRATION CATHETER 6
Type of Device
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key6120752
MDR Text Key60659344
Report Number3005168196-2016-01660
Device Sequence Number1
Product Code DXE
UDI-Device Identifier00814548016245
UDI-Public00814548016245
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K160533
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date07/20/2019
Device Catalogue NumberCAT6
Device Lot NumberF70718
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/24/2016
Initial Date FDA Received11/22/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/20/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-