Catalog Number 227445500 |
Device Problem
Break (1069)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/19/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
|
|
Event Description
|
Drill bit tips broke while drilling acetabular screws.
|
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
|
|
Event Description
|
Drill bit tips broke while drilling acetabular screws.
|
|
Manufacturer Narrative
|
Examination of the returned drills confirmed the complaint; the tips broke off.A complaint database search has identified a trend for this failure within the 2274 quickset drill family.Previous investigations have determined that use error is the likely root cause.As a result of trending health hazard evaluation, dva-106292-hhe was conducted.The investigation did not establish the need for corrective action.No evidence was found of product or design error as a contributing factor.Complaints will be monitored under sep 419 post market surveillance.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
|
|