• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - IRVINE FOLFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - IRVINE FOLFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C4063K
Device Problems Fluid/Blood Leak (1250); Material Perforation (2205)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/22/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a large volume folfusor leaked during infusion.The reporter stated that the leak was from a ¿gash¿ in the tube about 0.5cm from the flow regulator.The device had been filled with tazocin.The reporter stated that this was noted approximately a day after the start of infusion.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The lot was manufactured february 27, 2016 - march 01, 2016.The device was received for evaluation.Visual inspection noted a segment on the tubing was partially cut.The reported condition was verified.The cause of the condition was determined to be damage during use.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE - IRVINE
irvine CA
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong ave.
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6125606
MDR Text Key60857036
Report Number1416980-2016-17618
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number2C4063K
Device Lot Number16B046
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/02/2016
Initial Date FDA Received11/23/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/20/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
TAZOCIN
-
-