Model Number MODEL 250
Device Problems Application Program Problem: Dose Calculation Error (1189); No Device Output (1435); Device Stops Intermittently (1599)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/02/2016
Event Type  malfunction  
Event Description
It was reported that a programming computer¿s is not responsive and a hard reset could not resolve the issue.The handheld works intermittently and does affect patients¿ dosing sessions.Review of manufacturing records confirmed all tests passed for the device prior to distribution.
 
Event Description
The suspect programming computer was received by the manufacturer.No flashcard with programming software and no battery was returned inside handheld.An analysis was performed on the returned handheld and malfunction was confirmed.During the analysis it was identified that the handheld was unable to advance past the screen alignment utility.The cause for the display anomaly is associated with resistance values being higher than expected in the touch screen circuitry.No further anomalies associated with the handheld performance were identified during the analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6127164
MDR Text Key61142283
Report Number1644487-2016-02727
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2016
Is the Reporter a Health Professional? No
Event Location Other
Initial Date Manufacturer Received 11/02/2016
Initial Date FDA Received11/25/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/14/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N