• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP-CHINA
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/23/2016
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A nurse reported that a system crashed during a procedure.The procedure was completed after rebooting the system.There was no patient harm.
 
Manufacturer Narrative
Additional information: the system was examined and the reported event was replicated.The company representative reseated and cleaned the random-access memory (ram).The system was tested and found to meet product specifications.A review of the customer¿s complaint history for the last 24 months did not show any previous complaints of this kind against the system.The system was manufactured on july 30, 2013.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to the ram.However, how or when the ram became nonconforming cannot be determined (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
rita lopez
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514846
MDR Report Key6131705
MDR Text Key61302897
Report Number2028159-2016-05435
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP-CHINA
Device Catalogue Number8065751536
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/22/2016
Initial Date FDA Received11/29/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/28/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/30/2013
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-