• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 802100
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 11/04/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The reported complaint was confirmed by the product surveillance technician (pst) as per laboratory report.The pst observed blank screen at boot up and inaccurate touch screen response.Replacement of customer inverter board and customer touch screen restored central control monitor (ccm) functionality.The device was sent to service to be brought to manufacturer's specifications before returning to clinical use.This complaint is related to medwatch #1828100-2016-00653.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
It was reported that during routine testing of the device at the service center, the senior quality assurance technician found that the central control monitor (ccm) has a defective touch screen.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
eileen dorsey
6200 jackson road
ann arbor, MI 48103
7347416074
MDR Report Key6132017
MDR Text Key61403583
Report Number1828100-2016-00767
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number802100
Device Catalogue Number802100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/04/2016
Initial Date FDA Received11/29/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/27/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/29/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction Number1828100-06/05/2007-005-C
Patient Sequence Number1
-
-