Catalog Number 10379675 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/15/2016 |
Event Type
No Answer Provided
|
Manufacturer Narrative
|
The customer reported that they were exchanging the unit for a new instrument and sending the defective instrument in for investigation.The investigation of the instrument determined that the root cause was a failure of d16 on the main pcb.D16 is the overvoltage protector diode, which blows if the voltage applied to the instrument is too high.
|
|
Event Description
|
The customer reported that their clinitek status+, upon powering up, was smoking and produced a burning smell.There was no report of injury due to this event.
|
|