• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET ORTHOPEDICS COBALT HV BN CMT 40G-JAPAN; BONE CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET ORTHOPEDICS COBALT HV BN CMT 40G-JAPAN; BONE CEMENT Back to Search Results
Model Number N/A
Device Problem Incomplete or Missing Packaging (2312)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/31/2016
Event Type  malfunction  
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly or deviation.This product is manufactured by zimmer biomet (b)(4) and is not cleared or distributed in the u.S.However, this report is being submitted as zimmer biomet in (b)(4) manufactures a similar device in the united states under 510k number k150850.
 
Event Description
It was reported that during a procedure, that the liquid polymer was missing in the cement system package.Subsequently, another cement system was used to complete the procedure.
 
Manufacturer Narrative
This follow-up report is being filed to relay additional information, which was unknown at the time of the initial medwatch.
 
Manufacturer Narrative
This report is being filed to relay additional information.(b)(4).
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was not confirmed.Visual inspection of the returned product determined that the product has been opened and the only items contained within the packaging are the powder, the patient stickers, and the ifu.As the product was returned opened and the number of components issued to the order matches the quantity manufactured on the router, the reported event cannot be confirmed.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Review of the complaint history determined that no further action(s) are required.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBALT HV BN CMT 40G-JAPAN
Type of Device
BONE CEMENT
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6133198
MDR Text Key61087066
Report Number0001825034-2016-04903
Device Sequence Number1
Product Code MBB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PN/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 05/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date10/31/2017
Device Model NumberN/A
Device Catalogue Number402282J
Device Lot Number540390
Other Device ID NumberSEE H10 NARRATIVE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/31/2016
Initial Date FDA Received11/29/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Supplement Dates FDA Received02/08/2017
05/05/2017
05/08/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/31/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-