Catalog Number QD11 |
Device Problems
Vibration (1674); Naturally Worn (2988); Noise, Audible (3273)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was duplicated and confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|
Event Description
|
It was reported in the service order from (b)(6) that the attachment device made an abnormal noise and was vibrating.During pre-testing, it was observed that the device bearings were worn.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The date of manufacturer was documented as may 8, 2013 on the initial report.It has been updated as may 8, 2015.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|