• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS 11CM ANGLE ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS 11CM ANGLE ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT Back to Search Results
Catalog Number QD11
Device Problems Vibration (1674); Naturally Worn (2988); Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was duplicated and confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported in the service order from (b)(6) that the attachment device made an abnormal noise and was vibrating.During pre-testing, it was observed that the device bearings were worn.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The date of manufacturer was documented as may 8, 2013 on the initial report.It has been updated as may 8, 2015.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11CM ANGLE ATTACHMENT
Type of Device
MOTOR, DRILL, ELECTRIC - ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key6134624
MDR Text Key61440997
Report Number1045834-2016-13297
Device Sequence Number1
Product Code HBC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K011444
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberQD11
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/14/2016
Initial Date FDA Received11/30/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/16/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/08/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-