• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD ADULT NASAL INTERFACE; CAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD ADULT NASAL INTERFACE; CAT Back to Search Results
Model Number OPT946
Device Problem Device Issue (2379)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).We are in the process of obtaining a complaint cannula for evaluation.We will provide a follow up report upon completion of our investigation.
 
Event Description
A hospital in (b)(6) reported that the tubing of the opt946 nasal cannula is disconnecting from the distal connector that attaches to the breathing circuit.No specific incident details were reported.
 
Manufacturer Narrative
(b)(4).The opt946 interface is used to deliver humidified oxygen to patients and consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Method: no complaint cannula was received, however 86 sealed, unused cannulae were returned to fisher & paykel healthcare for evaluation.The unused samples were visually inspected and checked for defects.Results: visual inspection of the unused cannulae revealed no visible defects.Five of the unused cannulae were removed from the packaging and more closely inspected.No visible damage was found and it was not possible to detach the tubing from the swivel connectors.The swivels were found to be operational.Conclusion: we were unable to confirm the complaint as reported by the customer as no defects were found with the returned samples.All optiflow interfaces are inspected during production for visual defects including cracks, tears, inclusions, discolouration and stretching or deformation.Any product that fails the visual inspection is disposed of.The setup instructions in the user instructions which accompany the opt900 series nasal cannula include the following steps: - ensure head strap clip is attached, to prevent cannula from being pulled out of the nares.- cannula can become unattached if not used with the head strap clip.- attach tubing clip to clothing/bedding to prevent cannula from pulling off face.The user instructions also contain the following warnings/cautions: - do not crush or stretch tube, to prevent loss of therapy.- failure to use the set-up described above can compromise performance and affect patient safety.
 
Event Description
A hospital in (b)(6) reported that the tubing of the opt946 nasal cannula is disconnecting from the distal connector that attaches to the breathing circuit.No specific incident details were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT NASAL INTERFACE
Type of Device
CAT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
raymond yan
173 technology drive
suite 100
irvine, CA 92618
9194534000
MDR Report Key6137286
MDR Text Key61191865
Report Number9611451-2016-00818
Device Sequence Number1
Product Code CAT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Respiratory Therapist
Type of Report Initial,Followup
Report Date 11/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPT946
Device Catalogue NumberOPT946
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/01/2016
Initial Date FDA Received11/30/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/11/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-