• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event Description
On 11/08/2016, the reporter contacted animas alleging that the display screen was blank.There was no indication that the device caused or contributed to an adverse event.This complaint is being reported because the issue may result in an inability to use the product which may lead to long term cessation of delivery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key6139464
MDR Text Key61496828
Report Number2531779-2016-32641
Device Sequence Number1
Product Code MDS
UDI-Device Identifier10840406100198
UDI-Public0110840406100198
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 11/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Other Device ID Number1-MCBC-3762
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Device Age8 MO
Initial Date Manufacturer Received 11/08/2016
Initial Date FDA Received12/01/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/24/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age12 YR
-
-