Northgate technologies inc.Is the original equipment manufacturer (oem) of the 50l insufflator, which is the subject of this reportable event.Novadaq technologies inc.Is the re-packager/re-labeler of this 50l insufflator.Novadaq's name and branding is present on the insufflator's packaging and labeling; that is, the device is labeled as the "novadaq 50l insufflator".Since this device malfunction/product problem occurred with the novadaq-branded insufflator, novadaq technologies inc.Is therefore the party submitting this report.
|
A 50l insufflator (manufactured by northgate technologies) was being used with novadaq tubing and novadaq pinpoint system for an entire rt hemicolectomy case.The monitor was set at high flow: 25 and pressure: 15.About an hour and a half into the case, the patient's abdomen became over-inflated.The actual pressure on the insufflator monitor read 35 and was rising up to 40.The patient's co2 level was rising past 60 when the physician decided to end the laparoscopic portion of the procedure early due to the insufflator not functioning properly.After the open portion of the procedure, the surgeon decided to go back in laparoscopically, despite the clinical representative's recommendation to not use the insufflator again.The surgeon used the insufflator for a total of about 5 minutes and there were no issues.
|