STRYKER ORTHOPAEDICS-MAHWAH DETACHABLE FLEX SHAFT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
|
Back to Search Results |
|
Catalog Number 2107-2200 |
Device Problems
Break (1069); Fracture (1260); Material Integrity Problem (2978)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/07/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of the investigation.
|
|
Event Description
|
During surgery, the drill shaft broke.
|
|
Manufacturer Narrative
|
An event regarding an alleged fracture involving a trident driver shaft was reported.The event was confirmed.Method & results: device evaluation and results: device was returned in used condition and broken in 2 pieces.The flexible shaft had separated near the point of juncture with the drive fitting.At the point of separation, individual flexible shaft cables were twisted and frayed.Medical records received and evaluation: no medical records were received for review with a clinical consultant.Device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.Complaint history review: there has been no other event for the lot referenced.Conclusions: the reported event was confirmed.The flexible shaft had separated near the point of juncture with the drive fitting.At the point of separation, individual flexible shaft cables were twisted and frayed.Examination of the returned device with material analysis engineer indicated it fractured due to overload conditions.
|
|
Event Description
|
During surgery, the drill shaft broke.
|
|