STAAR SURGICAL COMPANY ICL (IMPLANTABLE COLLAMER LENS); INTRAOCULAR LENS Back to Search Results
Model Number MICL13.2
Device Problem Entrapment of Device (1212)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/31/2016
Event Type  malfunction  
Manufacturer Narrative
Event problem and evaluation codes: results codes: visual inspection of the returned product found the lens stuck inside an sfc-45 fp cartridge.The cartridge tip was split and a foam tip plunger was stuck inside.There was evidence of dry surgical residue.Conclusions code: based on the complaint history, work order search and the evaluation of the returned product, a specific root cause of the event could not be determined.(b)(4).
 
Event Description
The reporter stated a 13.2mm micl13.2 implantable collamer lens, -9.0 diopter, became stuck in the loader.There was no patient contact and the backup lens was used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICL (IMPLANTABLE COLLAMER LENS)
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
michelle andres
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key6140268
MDR Text Key61298034
Report Number2023826-2016-01710
Device Sequence Number1
Product Code MTA
UDI-Device Identifier00841542103275
UDI-Public(01)00841542103275(17)180430
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 11/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date04/30/2018
Device Model NumberMICL13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/04/2016
Initial Date FDA Received12/01/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/09/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARTRIDGE MODEL SFC-45 FP - LOT NUMBER 1330315; FOAM TIP PLUNGER MODEL FTP - LOT NUMBER UNK; INJECTOR MODEL MSI-PF - LOT NUMBER UNK
Patient Age33 YR
-
-