Model Number 383400 |
Device Problems
No Display/Image (1183); Failure to Power Up (1476)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Submit date: 12/02/2016.An investigation is currently underway.Upon completion, the results will be forwarded.
|
|
Event Description
|
It was reported to covidien on (b)(6) 2016 that an issue occurred with a feeding pump.Customer reports that while they were flushing the pump, the screen went blank and then the pump would not turn on.No patient involvement.
|
|
Manufacturer Narrative
|
An evaluation of the kangaroo pump was performed for the reported condition that the screen went blank and then the pump would not turn on.The unit was triaged and the customer¿s reported condition was confirmed.The potential root causes are the use of an unauthorized power adapter by the user and a possibly defective component.A review of the device history record by (b)(6) on 2017-oct-17 shows that this unit was manufactured in 2013 and was released meeting all manufacturing specifications.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
|
|