Model Number FHC-A202
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The customer's observation was not replicated in-house with retention products.Retention products were tested with 20miu/ml hcg and 3 high levels of hcg urine controls.Retention products showed hcg positive results at read time and met qc specifications.No false negative results were obtained.Manufacturing batch record review did not uncover any abnormalities.Root cause could not be determined based on the information provided by customer.Based on the information available, there is no indication of a product deficiency.No corrective action is required at this time.
 
Event Description
Report received of two false negative hcg results for emergency department patients using cardinal health rapid test hcg combo 30t.Customer questioning why a negative result would eventually become positive after read time.Customer did not provide patient specific information.Customer reported they used 3 drops, read time 3 mins received negative results (then read again greater than 4mins, maybe as long as 30mins later, and result was positive).No reported adverse patient sequela.No additional information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDINAL HEALTH RAPID TEST HCG COMBO 30T
Type of Device
HCG PREGNANCY TEST
Manufacturer Contact
ya-ling king
9975 summers ridge road
san diego, CA 92121
8588052084
MDR Report Key6142491
MDR Text Key61402453
Report Number2027969-2016-00699
Device Sequence Number1
Product Code JHI
Combination Product (y/n)N
PMA/PMN Number
K062361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Type of Report Initial
Report Date 11/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFHC-A202
Device Lot NumberHCG6060066
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/09/2016
Initial Date FDA Received12/02/2016
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-