Model Number 103
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Inadequate Pain Relief (2388)
Event Date 01/01/2016
Event Type  Injury  
Event Description
It was initially reported that the patient underwent generator replacement prophylactically on (b)(6) 2016.Additional surgical report was received on 11/10/2016 indicated that the patient is having electric sensation and pain into his left shoulder and decreased effectiveness of vns.The explanted generator was received on 11/08/2016.Analysis is underway but has not been completed to date.No additional relevant information was received.
 
Event Description
Analysis of the generator was completed and proper functionality of the pulse generator in its ability to provide appropriate programmed output currents was successfully verified in the lab.In addition, the septum was not cored, thus eliminating the possibility of a potential unintended electrical current path through body fluids.In the lab, the device output signal was monitored for more than 24-hrs, while the pulse generator was placed in a simulated body temperature environment.Results showed no signs of variation in the pulse generator¿s output signal and demonstrated that the device provided the expected level of output current for the entire monitoring period.The pulse generator diagnostics were as expected for the programmed parameters.In addition, a comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.The battery, 2.706 volts, shows an ifi=yes condition.There were no performance or any other type of adverse conditions found with the pulse generator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 103
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6143441
MDR Text Key61401204
Report Number1644487-2016-02785
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date12/31/2012
Device Model Number103
Device Lot Number2913
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2016
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 11/10/2016
Initial Date FDA Received12/02/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/13/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/10/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age60 YR
-
-