A device history record ( dhr) review of the reported lot indicates that the order was built to specification.Seven wet, used fluidics management system (fms) cassettes were returned for this complaint.The samples were visually inspected and no obvious defects were found; however, it was noted that the drain bags were cut off underneath the drain bag tape.The samples were functionally tested and samples 1 through 6 passed testing.Sample 7, failed to prime and generated a system message.No fluid flowed from the aspiration line to the cassette.The aspiration line was check for occlusions.The tubing was cut 27 inches from the luer insertion.Neither air pressure from the mensor, nor water from a syringe could purge the occluded silicone-like surgical debris.An aspiration manifold from the lab stock was used to complete functional testing on sample 7.The sample passed functional testing with the lab stock aspiration manifold.The root cause of this customer's complaint could not be established as six of the seven returned samples met specifications and the seventh sample initially failed priming due to occluded surgical debris, but passed with a lab stock aspiration manifold.It is possible that the surgical debris hardened in the tubing during transit.The sample passed functional testing with a lab stock manifold free of occluded surgical debris.An internal investigation has been opened to address reports of a similar nature.(b)(4).
|