• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY ARCADIS VARIC; SYSTEM, X-RAY, FLUOROSCOPIC, IMAGE-INTENSIFIED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY ARCADIS VARIC; SYSTEM, X-RAY, FLUOROSCOPIC, IMAGE-INTENSIFIED Back to Search Results
Model Number 10143406
Device Problems Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/02/2016
Event Type  malfunction  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.(b)(6).
 
Event Description
It was reported to siemens that a malfunction occurred while operating the arcadis varic system.The system was in service for approximately 5 minutes using fluoro.The customer reported error asp_msy_5004.A restart of the system was attempted without success and the procedure was terminated.We are unaware of any impact to the state of health of the patient involved.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The logfiles showed the error "asp_msy_5004".When this error happens the message: "generator: intermediate circuit voltage failure.If the failure occurs frequently, inform service." will be displayed.A complete system shut down solves this error.The problem was resolved after performing a generator adjustment.This event does not exceed the threshold in regards to the defect acceptance level and a general field corrective action is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCADIS VARIC
Type of Device
SYSTEM, X-RAY, FLUOROSCOPIC, IMAGE-INTENSIFIED
Manufacturer (Section D)
SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY
siemensstrasse 1
forcheim, 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY
siemensstrasse 1
forcheim, 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
65-1a
malvern, PA 19355
6104486461
MDR Report Key6144662
MDR Text Key61711913
Report Number2240869-2016-06311
Device Sequence Number1
Product Code JAA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K040066
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 11/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number10143406
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/30/2016
Initial Date FDA Received12/05/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/23/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-