BIOMET ORTHOPEDICS COMPREHENSIVE SEGMENTAL REVISION SYSTEM DISTAL BODY WITH SCREW-RIGHT 70MM; PROSTHESIS, SHOULDER Back to Search Results
Model Number N/A
Device Problem Device Operational Issue (2914)
Patient Problem No Information (3190)
Event Date 11/08/2016
Event Type  Injury  
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show the lot released with no recorded anomaly or deviation.(b)(4).Medical product - disc condyle kit w/ hexalobula, catalog#: 114700 lot 721800; compr srs mod stem - 8x100mm, catalog#: 211236 lot 263930; compr srs small flange, catalog#: 211269 lot 652190; disc ulna 3x115mm rt w/brng c, catalog#: 114817 lot ni.Therapy date - (b)(6) 2016.
 
Event Description
During an elbow revision of competitor product, the distal humerus component would not fit with the condyle kit.A high speed burr was utilized to remove metal from the component to facilitate assembly.The medial condyle screw also remained proud and was tight.This event caused a 75 minute delay.
 
Manufacturer Narrative
This follow up report is being filed to relay additional information.No device or photos were received; therefore the condition of the device is unknown.Review of device history records revealed no deviations or anomalies.The instructions for use indicate the devices used are compatible.Complaint history search found no additional complaint for this device.A definite root cause cannot be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPREHENSIVE SEGMENTAL REVISION SYSTEM DISTAL BODY WITH SCREW-RIGHT 70MM
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6145731
MDR Text Key61458466
Report Number0001825034-2016-05032
Device Sequence Number1
Product Code KWT
Combination Product (y/n)N
Reporter Country CodeFI
PMA/PMN Number
PK111746
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number211255
Device Lot Number688560
Other Device ID NumberSEE NARRATIVE IN H10
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/08/2016
Initial Date FDA Received12/05/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/25/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/29/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE NARRATIVE IN H10
Patient Outcome(s) Required Intervention;
Patient Age78 YR
Patient Weight63
-
-