The mdr with manufacturing number 1644487-2016-02574 captures the reported high impedance, change in seizure frequency, and change in seizure pattern.During the revision surgery on (b)(6) 2016 the surgeon opened the generator pocket first and noticed the plastic header of the 102r was completely disconnected from the generator canister.The generator was received for analysis on 12/05/2016.Product analysis is underway but has not been completed to date.
|
Product analysis was completed on the explanted generator and found that the header was completely detached from the can.There was no noted corrosion or body fluids observed in the detachment area, which means that the detachment occurred recently.Adequate adhesive residue was seen on the can in the header area.Visual observations did not show any tool marks on the can or header, which would indicate that the explant procedure was not the cause of the header detachment.Product analysis found that the header became detached during the implant period and not during the explant surgery.The generator was then altered to allow lead wires to connect to the circuit board and test the internal circuitry of the generator.There were no functional anomalies identified with the circuitry.Other than the detachment of the header from the can no anomalies were observed.
|