CAREFUSION, INC 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY
|
Back to Search Results |
|
Model Number 3100A |
Device Problems
Calibration Problem (2890); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/07/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).At this time, carefusion has not received the suspect device/component for evaluation.In the event that the device is received for evaluation or additional information is received, a follow-up report will be submitted.
|
|
Event Description
|
The customer reported that the users are having intermittent issues with the patient circuit calibration.Reportedly, sometimes the unit is reading 41 cm and sometimes the unit is only reading 10 cm.There was no patient involvement associated with the reported issue.
|
|
Manufacturer Narrative
|
Results of investigation: a carefusion field service representative (fsr) evaluated the device onsite.The fsr found that when the settings were set under 10 cmh2o, the unit would stop.The fsr calibrated the pneumatics, driver displacement indicator (ddi), and the airway pressure transducer.The fsr ran the operational verification procedure (ovp) and the unit meets manufacturer specifications.
|
|