CYBERONICS - HOUSTON PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problems Break (1069); Detachment Of Device Component (1104); Component Falling (1105); Loose or Intermittent Connection (1371); Communication or Transmission Problem (2896)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/11/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported by a company representative that the physician's tablet usb port is broken.The cable connected to the tablet felt really loose so the representative tried removing the cable and putting the cable back in.When the cable was inserted back into the tablet, the cable wouldn¿t stay in the tablet.Per the company representative, the usb port was loose within the tablet and fell off.The physician had backup equipment and so they were still able to interrogate the patient and no patient was affected.A replacement tablet was provided to the physician.The suspect tablet was received on 11/22/2016.An analysis was performed on the returned tablet and the reported allegation was verified.During the analysis it was identified that the usb port was damaged.As a result, the tablet was unable to establish communication.No further anomalies were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6150809
MDR Text Key62144699
Report Number1644487-2016-02807
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 11/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 11.0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2016
Is the Reporter a Health Professional? No
Event Location Other
Initial Date Manufacturer Received 11/11/2016
Initial Date FDA Received12/07/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-