• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD. 8010379 TOTAL ASR FEM IMP SIZE 53; HIP FEMORAL HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD. 8010379 TOTAL ASR FEM IMP SIZE 53; HIP FEMORAL HEAD Back to Search Results
Catalog Number 999804653
Device Problems Appropriate Term/Code Not Available (3191); Noise, Audible (3273)
Patient Problem No Information (3190)
Event Date 08/04/2014
Event Type  Injury  
Manufacturer Narrative
No 510(k) number provided because this implant is sold internationally with different indications for use; it is currently sold in the us under a different part number.The correction/removal reporting number listed applies to the corresponding product code sold domestically.The asr platform was voluntarily recalled from the market in august 2010, and the asr product codes are now considered inactive.Further investigation of this individual incident will not be undertaken, as there is an ongoing investigation regarding the root cause(s) and/or corrective actions.(b)(4).Depuy considers the investigation closed at this time.Should the product and/or additional information be received, the investigation will be re-opened.
 
Event Description
Patient underwent a revision to address elevated metal ions levels and noise.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> the asr platform was voluntarily recalled from the market in august 2010, and the asr product codes are now considered inactive.Further investigation of this individual incident will not be undertaken.The investigation regarding the root cause(s) and/or corrective actions was conducted under mdd capa- (b)(4).Ongoing post market surveillance is conducted per our procedures for this product.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Asr revision.Asr resurfacing, left.Reasons for revision : noise / increased serum concentrations from metal ions.Update - alert date 25 nov 2016.Legal claim received - (b)(6).Update ad 17 nov 2018: (b)(4) has been re-opened under (b)(4) due to receipt of claimsuite alert.Corrected lot number for the acetabular cup.Doi: (b)(6) 2006; dor:(b)(6) 2014; left hip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOTAL ASR FEM IMP SIZE 53
Type of Device
HIP FEMORAL HEAD
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD. 8010379
st. anthony's rd
leeds LS11 8DT
UK  LS11 8DT
Manufacturer (Section G)
LEEDS MFG & MATERIAL WAREHOUSE
st anthonys road
leeds LS11 8DT
UK   LS11 8DT
Manufacturer Contact
kara ditty-bovard
1210 ward avenue
west chester, PA 19380
6103142063
MDR Report Key6154965
MDR Text Key61806450
Report Number1818910-2016-33093
Device Sequence Number1
Product Code KXA
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,other
Reporter Occupation Patient
Remedial Action Recall
Type of Report Initial,Followup
Report Date 11/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/23/2010
Device Catalogue Number999804653
Device Lot Number1949445
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/23/2016
Initial Date FDA Received12/08/2016
Supplement Dates Manufacturer Received04/19/2019
Supplement Dates FDA Received04/19/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-