• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO REVO MRI SURESCAN; PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO REVO MRI SURESCAN; PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE Back to Search Results
Model Number RVDR01
Device Problems Unable to Obtain Readings (1516); Reset Problem (3019); Missing Test Results (3267)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/25/2016
Event Type  Injury  
Manufacturer Narrative
Concomitant products: 5086mri52 lead implanted: (b)(6) 2011.
 
Event Description
It was reported that the patient was previously admitted and was back in the emergency department.The device experienced an electrical reset, with battery voltage not being measured and invalid data seen.Atrial undersensing was also noted.The device was interrogated thus clearing the reset, and both the device and atrial lead remain in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVO MRI SURESCAN
Type of Device
PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6161525
MDR Text Key62104285
Report Number3004209178-2016-25764
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P090013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 10/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/28/2012
Device Model NumberRVDR01
Device Catalogue NumberRVDR01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/25/2016
Initial Date FDA Received12/09/2016
Supplement Dates Manufacturer Received10/25/2016
Supplement Dates FDA Received09/27/2017
Date Device Manufactured05/09/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
5086MRI45 LEAD
Patient Outcome(s) Required Intervention;
Patient Age87 YR
-
-