• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TRUEPASS SUTURE PASSER SELF-CAPTURE; ACCESSORIES,ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TRUEPASS SUTURE PASSER SELF-CAPTURE; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Catalog Number 72203791
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/07/2016
Event Type  malfunction  
Event Description
It was reported that there is a small spring that forces the capture mechanism to close and holds the suture in place, this spring came out of its place.No patient injury or complications were reported.
 
Manufacturer Narrative
One truepass suture passer w/self-capture feature was returned for evaluation.Visual assessment of the returned device confirmed the torsional spring to be dislodged from the recessed grooves but remains attached to the top jaw.A root cause investigation has been opened to address this failure mode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEPASS SUTURE PASSER SELF-CAPTURE
Type of Device
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer Contact
james gonzales
7000 w. william cannon drive
austin, TX 78735
5123585706
MDR Report Key6166291
MDR Text Key62420095
Report Number1219602-2016-01360
Device Sequence Number1
Product Code NBH
Combination Product (y/n)N
Reporter Country CodeCY
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 11/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72203791
Device Lot Number27869
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/15/2016
Initial Date FDA Received12/12/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received01/19/2017
01/20/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age45 YR
-
-