ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP
Device Problem Air Leak (1008)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/16/2016
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A company representative reported that he observed air bubbles flowing from the infusion cannula into the patient's eye during two separate vitrectomy procedures on the same day.The product samples were not kept.There was no harm to the patients.Additional information has been requested but not received to date.
 
Manufacturer Narrative
As the customer did not retain the finished goods lot number, device history review and lot history could not be reviewed.The customer did not retain a sample for this complaint report therefore a visual inspection or functional testing could not be conducted.The root cause of the customer's complaint could not be established as a sample has not been received.Without a sample, it is not possible to isolate the root cause.Potential root cause could be related to a manufacturing related issue, however, this could not be confirmed with the information available.After a thorough investigation of this complaint, it has been determined that no action will be taken at this time as a sample was not returned.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
rita lopez
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514846
MDR Report Key6166435
MDR Text Key62435752
Report Number2028159-2016-05663
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeLH
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Other Device ID Number3.00.61
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/16/2016
Initial Date FDA Received12/12/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/01/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/11/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
COMBINED PLUS PACK, 25 GAUGE, 5000 CUTS PER MINUTE
Patient Outcome(s) Other;
-
-