• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KANGAROO EPUMP - NEW; ENTERAL FEEDING PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO EPUMP - NEW; ENTERAL FEEDING PUMP Back to Search Results
Model Number 382400
Device Problems No Device Output (1435); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Submit date: 12/13/2016.An investigation is currently under way.Upon completion the results will be forwarded.
 
Event Description
It was reported to covidien on (b)(6) 2016 that a customer had an issue with an enteral feeding pump.The customer states the unit displays an error 33.Upon triage on (b)(6) 2016 the service tech found the unit had no audio.
 
Manufacturer Narrative
An evaluation of the kangaroo epump was performed for the reported condition of, ¿service technician finding of no audio during triage¿.It was noted that the pump had liquid ingress, a capacitor was missing, a piece of the pcb was burnt, and there was a dislodged operational chip.A trend has been identified and a corrective action has been opened to address this issue.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO EPUMP - NEW
Type of Device
ENTERAL FEEDING PUMP
Manufacturer (Section D)
COVIDIEN
building # 10, 789 puxing road
shanghai
CH 
MDR Report Key6171778
MDR Text Key62613535
Report Number3006451981-2016-00681
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup,Followup
Report Date 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number382400
Device Catalogue Number382400
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/06/2016
Initial Date FDA Received12/13/2016
Supplement Dates Manufacturer Received12/06/2016
08/08/2018
Supplement Dates FDA Received08/06/2018
10/03/2018
Patient Sequence Number1
-
-