It was reported that the physician was experiencing intermittent communication issues with his programming system.Troubleshooting included confirming that the wand¿s 9v battery was functioning properly and it was confirmed that the tablet was not plugged into the wall outlet.The physician was eventually able to program however since the issues were intermittent it was noted to be difficult.A company representative later went to the physician¿s office and performed troubleshooting.It was then confirmed that the issue resolved after the serial cable was replaced.The suspect serial cable was discarded.The serial cable does not require return, as the failure mode is understood to be a failure of the serial cable associated with a disconnected wire connection.
|