Model Number 91496
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/22/2016
Event Type  malfunction  
Manufacturer Narrative
Onsite testing of the involved devices conducted by a spacelabs field service engineer confirmed that the equipment performed to specifications.The testing was witnessed by a facility staff member.The patient's historical waveforms and trend information were reviewed by a spacelabs lead software engineer.Findings show that a vrun alarm was generated for the episode.The device performed to specifications.The investigation is considered complete, and the issue closed.
 
Event Description
Spacelabs received a report that on (b)(6) 2016, at 01:30 an episode of ventricular tachycardia (vtach) did not alarm at the central monitor.No one was injured as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPACELABS ULTRAVIEW SL COMMAND MODULE
Type of Device
ULTRAVIEW SL MULTIPARAMETER MODULE
Manufacturer (Section D)
SPACELABS HEALTHCARE INC.
35301 se center st.
snoqualmie WA 98065
Manufacturer Contact
marc rivas
35301 se center st.
snoqualmie, WA 98065
4253635568
MDR Report Key6174071
MDR Text Key62688584
Report Number3010157426-2016-00196
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103142
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Health Professional
Type of Report Initial
Report Date 12/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number91496
Other Device ID Number2.05.12
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/22/2016
Initial Date FDA Received12/14/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
91390 ((B)(4)) 96102
-
-