MEDIVATORS DSD-201; AUTOMATED ENDOSCOPE REPROCESSOR Back to Search Results
Device Problem Maintenance Does Not Comply To Manufacturers Recommendations (2974)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/15/2016
Event Type  No Answer Provided  
Manufacturer Narrative
Medivators field service engineer discovered while performing a pm on a dsd-201 aer that the filters were not being changed out in accordance to the aer ifu.Improper filter and machine maintenance could lead to improper endoscope disinfection, thus potential patient harm.Medivators fse informed the customer the importance of filter replacement per machine maintenance guidelines and documenting appropriately.To date, there have been no reported patient illness or injury.This complaint will continue to be maintained within medivators complaint handling system.
 
Event Description
Medivators field service engineer discovered while performing a pm on a dsd-201 that the filters were not being changed out in accordance to the ifu.Improper filter and machine maintenance could lead to improper endoscope disinfection, thus potential patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DSD-201
Type of Device
AUTOMATED ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
MEDIVATORS
14605 28th ave n
minneapolis MN 55447
Manufacturer (Section G)
MEDIVATORS
14605 28th ave n
minneapolis MN 55447
Manufacturer Contact
megan dickey
14605 28th ave n
minneapolis, MN 55447
7635533327
MDR Report Key6175491
MDR Text Key62422688
Report Number2150060-2016-00051
Device Sequence Number1
Product Code FEB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K914145
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Medical Equipment Company Technician/Representative
Remedial Action Notification
Type of Report Initial
Report Date 12/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/15/2016
Initial Date FDA Received12/15/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/20/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-