• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY PROFESSIONAL 30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT - 10S STRAIGHT; SCALER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY PROFESSIONAL 30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT - 10S STRAIGHT; SCALER, ULTRASONIC Back to Search Results
Catalog Number 80395
Device Problems Overheating of Device (1437); Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
While no serious injury resulted in this event, there has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
While using a 30k fsi-10s insert, the insert was heating up; no injury resulted.
 
Manufacturer Narrative
Insert had ~95% obstruction due to rust.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT - 10S STRAIGHT
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY PROFESSIONAL
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY PROFESSIONAL
1301 smile way
york PA 17404
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key6182891
MDR Text Key63069327
Report Number2424472-2016-00090
Device Sequence Number1
Product Code ELC
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K052334
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 01/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number80395
Device Lot Number16113
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/16/2016
Initial Date FDA Received12/16/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/25/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-