|
Model Number EIS-CFR-25 |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/30/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been returned to olympus for evaluation.The cause of the reported event could not be determined; however, the most probably cause could be attributed to the operator's technique.The instruction manual states, "do not use an instrument that fails to meet the criteria stated in the labeling or that has been damaged.Damage may result in the loss of the entire ceramic tip or fragments of the ceramic tip.If there is evidence of charring, burn spots, chips or cracks in the ceramic tip or surrounding area, do not use.".
|
|
Event Description
|
Olympus was informed that during a transurethral resection of bladder tumor (turbt) procedure, a portion of the ceramic tip broke off inside the patient.The broken tip was retrieved from the patient.The procedure was completed using a different device.No patient injury.
|
|
Manufacturer Narrative
|
The device was returned to olympus for evaluation.The evaluation confirmed the reported device damage.There was approximately (b)(6) of the ceramic beak missing from the distal end.The device fragment was not returned to olympus.In addition, there were scratches noted on the distal end beak.Based on the investigation findings and similar reported events, the cause of the reported event is likely due to the operator¿s technique.The instruction manual for use states, ¿do not use an instrument that fails to meet the criteria stated in the labeling or that has been damaged.Damage may result in the loss of the entire ceramic tip or fragments of the ceramic tip.If there is evidence of charring, burn spots, chips or cracks in the ceramic tip or surrounding area, do not use.¿.
|
|
Search Alerts/Recalls
|
|
|