• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY ARTIS ZEE CEILING; SYSTEM, X-RAY, ANGIOGRAPHIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY ARTIS ZEE CEILING; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number 10094137
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/16/2016
Event Type  malfunction  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.(b)(6).
 
Event Description
It was reported to siemens that a malfunction occurred while operating the artis zee ceiling system.The customer reported that the skull support was broken during a patient procedure.We are unaware of any impact to the state of health of any patient involved.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The investigation results showed marks on the skull support from a collision with the system or a drop on the floor.Collision of the skull support leads to high stresses in the connections and to eventual permanent damage.The skull support was exchanged and the system was returned to clinical use.No further actions will be taken by the manufacturer at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEE CEILING
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY
siemensstrasse 1
forcheim, 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH
siemenstrasse 1
forcheim, 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
65-1a
malvern, PA 19355
6104486461
MDR Report Key6188847
MDR Text Key63198668
Report Number2240869-2016-06717
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K073290
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 12/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number10094137
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/15/2016
Initial Date FDA Received12/20/2016
Supplement Dates Manufacturer Received12/15/2016
Supplement Dates FDA Received11/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-