The customer runs esoteric testing on an advia centaur xp instrument (cxp1) and (b)(6) testing on another advia centaur xp instrument (cxp2).After the incident occurred the ccc disabled in centralink, (b)(6) testing on the cxp1 instrument, to stop the test from downloading to that instrument.The customer ran a test patient, and testing was performing as expected.The cause of the (b)(6) results reported out was due to a user error.The instrument is performing according to specifications.No further evaluation of the device is required.
|
The customer contacted the siemens customer care center (ccc).The customer stated they erroneously validated and reported discordant results on patient samples tested for (b)(6).The customer accepted "x- no result" results in the centralink data management system, then manually validated the results and uploaded a value of "(b)(6)" in the laboratory information system.The samples were repeated, and corrected results were issued to the physician(s).There are no reports of patient intervention or adverse health consequences due to the (b)(6) values reported.
|