MAUDE Adverse Event Report: SMITH & NEPHEW, INC. DRILL, FLEXIBLE, ENDOSCPC, CANN, 4.5MM; ACCESSORIES, ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. DRILL, FLEXIBLE, ENDOSCPC, CANN, 4.5MM; ACCESSORIES, ARTHROSCOPIC Back to Search Results
Catalog Number 72202971
Device Problem Computer Operating System Problem (2898)
Patient Problem Bone Fracture(s) (1870)
Event Date 11/23/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(6).
 
Event Description
It was reported that when passing the drill through the flexible wire, it broke through the bone before breaking the cortex.
 
Manufacturer Narrative
Examination was not possible, as the device was not returned.The investigation was limited to the information provided.The investigation could not draw any conclusions about the reported event with the clinical details provided.A review of the device history record was performed which confirmed no inconsistencies.No further investigation is warranted at this time.
 
Manufacturer Narrative
One 4.5 flexible endoscopic cannulated drill bit was returned for evaluation.Visual assessment of the drill confirmed the reported complaint of breakage.The drill shaft has broken where it transitions from the 20 ½ to 9 revolutions per inch at the lazer cut double helix.No material voids are present at the break area.Broken distal portion of the drill was not returned for examination making a root cause prohibitive.Further investigation is not warranted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL, FLEXIBLE, ENDOSCPC, CANN, 4.5MM
Type of Device
ACCESSORIES, ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer Contact
jim gonzales
7000 w. william cannon
austin, TX 78735
5123585706
MDR Report Key6190340
MDR Text Key63231804
Report Number1219602-2016-01390
Device Sequence Number1
Product Code NBH
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72202971
Device Lot Number50554567
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/23/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/24/2016
Initial Date FDA Received12/20/2016
Supplement Dates Manufacturer ReceivedNot provided
04/30/2018
Supplement Dates FDA Received02/22/2017
05/11/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N