• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) NAV THORACIC PROBE TIP; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) NAV THORACIC PROBE TIP; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9734680
Device Problems Bent (1059); Material Deformation (2976)
Patient Problem No Patient Involvement (2645)
Event Date 09/23/2016
Event Type  malfunction  
Manufacturer Narrative
The thoracic probe was returned to the manufacturer for analysis.Testing found that the tip of the probe was severely twisted and that the back end of the probe had impact marks.This indicated a mechanical failure due to a bent instrument tip.
 
Event Description
A medtronic representative reported that, while in a spinal fusion, that the tip of a thoracic probe was bent.The site was able to switch to another probe and continue with the procedure.There was no reported delay to the procedure due to this issue.There was no impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAV THORACIC PROBE TIP
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
826 coal creek circle
louisville, CO 80027
MDR Report Key6191279
MDR Text Key62916973
Report Number1723170-2016-05828
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00613994870063
UDI-Public00613994870063
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K124004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 12/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9734680
Device Lot Number150226
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/23/2016
Initial Date FDA Received12/20/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/26/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-