The customer returned the meter and test strips.The returned meter and strips were tested in comparison to a retention meter & masterlot strips.Two human blood samples from warfarin donors and internal reference meters were used.Donor #1 hct: 42%, donor #2 hct: 47%.Donor #1 master lot: 2.4 inr, donor #1 customer's strips and customer's meter: 2.5 inr.Donor #2 master lot: 3.8 inr, donor #2 customer's strips and customer's meter: 3.8 inr.All inr values were within the specified maximum difference between measurements.No error messages occurred.The returned material and the retention material meet the specification.Relevant retention test strips (lot 128043-21) were tested in comparison with the current master lot coaguchek xs pt test strip (lot 124158-80).For this purpose two human blood samples from marcumar donors and two internal reference meters were used.No error messages occurred.Retention material was acceptable.
|